• Sy-2801f Low-Temperature Penetrometer Laboratory Instrument
  • Sy-2801f Low-Temperature Penetrometer Laboratory Instrument
  • Sy-2801f Low-Temperature Penetrometer Laboratory Instrument

Sy-2801f Low-Temperature Penetrometer Laboratory Instrument

After-sales Service: 24/Hrs Reply
Warranty: 1 Year, 1 Year
Customized: Customized, Customized
Structure: Portable, Samyon, Portable
After-Sales Service: 24/Hrs Reply
Measuring Range: (0~600) Penetration
Customization:
Diamond Member Since 2015

Suppliers with verified business licenses

Rating: 5.0/5
Manufacturer/Factory, Trading Company

Basic Info.

Model NO.
SY-2801F
Timing Control
5s, 8s, 10s, 12s, 30s, 60s Can Be Selected.
Resolution
0.01mm
Penetration Precision
+/-1 Penetration
Transport Package
Carton Case
Specification
SGS
Trademark
SAMYON
Origin
Cn
HS Code
903180
Production Capacity
100sets Per Month

Product Description

Summary
This instrument is designed and made as per T0604-2011 Penetration Test for Bituminous Materials in JTG E20-2011 Specification and Test Methods of Bitumen and Bituminous Mixture for Highway Engineering and ASTM D5 Standard Test Method for Penetration of Bituminous Materials. It is used to determine the penetration of pavement petroleum asphalt, modified asphalt, liquid petroleum asphalt and emulsified asphalt. It is also suitable to test solid particle, powder, colloid and raw-food materials such as cheese, glycine, butter, cream and leavening. It can be widely used in food industry, highway engineering and other industrial fields.
 
I.Main technical features
1. It adopts digital display to show the penetration number and two grades of timing control.
2. Specially designed low-temperature circulatory water bath can supply the constant temperature water into glass bath on the penetrometer. It creates a constant low temperature which can be preset for the specimen.
 
II. Main technical specifications                                                      
1. Circulatory water bath
(1)Power supply:   AC20V±10%, 50Hz
(2)Bath volume:   8.5L
(3)Temperature range:   (5.0~100.0)ºC
(4)Temperature accuracy:   ±0.1ºC
(5)Circulatory water quantity:    5L/min
(6)Pump lift:   1m
(7)Maximum power consumption:   1000W
(8)Dimension:   420mm×380mm×900mm
2. Penetrometer
(1)Timing controller:   5s or 60s can be selected. Bias within±0.1s
(2)Needle holder:    High accuracy digital display. Easy to use and observe. It can be moved to any position.
(3)Resolution:   0.01m
(4)Penetration accuracy:   ±1 penetration
(5)Maximum scale:   600 penetration
(6)Standard needle:   50mm, 2.5g±0.05g
(7)Dimension:   280mm×350mm×700mm
Sy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory InstrumentSy-2801f Low-Temperature Penetrometer Laboratory Instrument