Packaging Tightness Tester
MFT-900
Introduction
The instrument adopts the differential pressure comparison method of the measured object and the reference object and can use the high sensitivity of the differential pressure sensor to detect very small leaks. During testing the workpiece is put into a sealed container for pressure or vacuum. If the workpiece has a small leak the pressure in the sealed container will change to detect a small leak. The gas will fill the work piece cavity during inflation. A large leak was detected by the pressure method.
Performance characteristics
Meet the ASTM F2338, USP1207.1 standard
Semi-automatic detection suitable for small batch and multi-variety testing
The leak rate can be automatically converted into defect aperture um
The test result database is stored to facilitate quality management
Using compressed air to detect low cost
Performance parameters
Measurement method |
Differential pressure airtight leak detector |
Differential pressure sensor |
±2kPa error ≤ ±0.5% F.S. |
Differential pressure resolution |
0.1Pa |
Test pressure sensor |
Gauge pressure sensor error≤±1% F.S. |
Minimum detectable leak rate |
1x10-5Pa.m³/s (100kPa test pressure is equivalent to the defect aperture of the ampoule bottle 1µm |
Stand-alone internal volume |
≤4mL |
Display method |
10 inch touch screen LCD display (24-bit color) effective display size |
temperature range |
Use: 0-40°C Storage: -5-45°C <85% RH (no condensation) |
Voltage |
AC100V-240V 50/60Hz power ≤200W |
Weight |
Host: about 30kg pressure control box: about 10kg |
Audit trail |
≥5 years storage multiple event logs can be queried according to time |
Authority management |
Login with username and password level 4 authority |
Packaging Leak Tester MFT-900
It is suitable for sealing integrity testing of pharmaceutical packaging to prevent moisture, oxygen, and microorganisms from contaminating products. Medicines: bottled, bagged, boxed, ampoules, vials, cartridges, prefilled needles (PFS), (BFS), (FFS), etc.
Spray cans: plunger spray cans; bag-lined spray cans; "energy jacket" spray cans; flexible tube spray cans.
1. Features:
Comply with USP <1207>, ASTM F2338 standards.
Semi-automatic detection, suitable for small batch and multi-variety testing.
Non-destructive non-destructive testing, high accuracy, repeatability, sensitivity.
The instrument is used for vacuum pressure difference detection.
The leak rate can be automatically converted into defect aperture μm.
Database storage of test results for easy quality management.
Touch-type man-machine interface, simple and quick operation: after setting/selecting the test program, only need to manually put in/take out the test sample.
2. Advantage function:
Automatically test the flow rate and change the aperture size during the whole process.
Automatic leak rate calibration function.
Equipped with standard leaks (standard positive bottles, with third-party certification).
Four-level user authority management.
With audit trail function.
Split design, the test chamber is located above the host, and various test chambers can be provided according to different product types.
We also provide users with supporting services related to tightness testing, including positive bottle production, standard leak rate/annual leak verification, new sample mold customization, sample methodological parameter development and verification, etc.
The test cavity is customized according to customer needs to ensure that the test cavity is fully matched with the customer's product, and rapid and sensitive testing.
3. Specifications:
Differential pressure sensor |
±2kpa, error≤0.5%F.S. |
Differential pressure resolution |
0.1pa |
Test pressure sensor |
Error≤±1%F.S. |
Test pressure range |
-0.1~0.2MPa/-0.1~0.7MPa/-0.1~1.0MPa |
Minimum detectable aperture |
2μm
|
Display method |
10-inch touch color touch screen |
Voltage |
AC100V-240V, 50/60Hz, power ≤200W |
Weight |
Host: about 30kg |
Audit trail |
≥5 years of storage Multiple event logs, which can be queried according to time |
Authority management |
Login with username and password, level 4 authority |
Print function |
External micro printer |
History record |
≥5 years storage |
Data backup |
Support U disk to export data |
Signal output |
(4-20) mA, RS485, 12V alarm output |
Interface |
Ethernet USB RS485 |