• Sy-3536 Cleveland Open-Cup Flash Point Tester
  • Sy-3536 Cleveland Open-Cup Flash Point Tester
  • Sy-3536 Cleveland Open-Cup Flash Point Tester
  • Sy-3536 Cleveland Open-Cup Flash Point Tester
  • Sy-3536 Cleveland Open-Cup Flash Point Tester
  • Sy-3536 Cleveland Open-Cup Flash Point Tester

Sy-3536 Cleveland Open-Cup Flash Point Tester

After-sales Service: 24/Hrs Reply
Warranty: 1 Year
Customized: Customized
Structure: Portable
Power Supply: AC(220±10%)V, 50Hz
Heating Device: Quartz Tube Furnace Heating
Customization:

Contact Supplier

Diamond Member Since 2015

Suppliers with verified business licenses

Rating: 5.0/5
Manufacturer/Factory, Trading Company

Basic Info.

Model NO.
SY-3536
Test Flame Applicator
It Applies The Test Flame Automatically.
Thermometer
(-6~400)Oc
Ambient Temperature
(-10~50)Oc
Transport Package
Carton Case
Specification
SGS
Trademark
SAMYON
Origin
Cn
HS Code
903180
Production Capacity
100sets Per Month

Product Description

Sy-3536 Cleveland Open-Cup Flash Point Tester

Summary

This instrument is designed and made as per standards GB/T3536-2008 Petroleum Products - Determination of Flash and Fire Points - Cleveland open cup method and ASTM D 92 Standard Test Method for Flash and Fire Points by Cleveland Open Cup Tester. It is applicable to all petroleum products with flash points above 79ºC and below 400ºC except fuel oils.
 
I. Main technical features
1. It adopts special heating furnace to ensure the safety of test. The heating power is continuously adjustable.It can meet requirements of test.
2. The instrument adopts desktop structure.The stainless steel working table and furnace shell make it beautiful and elegant.
3. The instrument is fully self-contained complete.The operator can do determination as long as connecting with coal gas or other civil gas.
4. It equipped with flame extinguishing cover which accord with requirements of test.
 
II. Main technical specifications
1. Power supply  AC(220±10%)V, 50Hz.
2. Heating device Quartz tube furnace heating, no naked fire, explosion prevented.The power is adjustable from 0W to 600W.
3. Test flame applicator It applies the test flame automatically.
4.Thermometer (-6~400)ºC. It is the same as thermometer ASTM 11C.
5. Igniting device (1) Ignition source: coal gas(or civil gas,the same below)
(2) Flame diameter is 3.2mm~4.8mm
6. Ambient temperature (-1050)ºC
7. Relative humidity ≤85%
8. Maximum power consumption  650W
9. Dimension 350mm×290mm×350mm (thermometer is not included)
 
 
Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester Sy-3536 Cleveland Open-Cup Flash Point Tester
 